CANNON-Ubbelohde Dilution Viscometers
Catalog Number | Size | Approximate Constant (mm²/s²) | Viscosity Range (mm²/s)† | ||
---|---|---|---|---|---|
Uncalibrated | Calibrated | Low | High | ||
9722-L50 | 9722-M50 | 25 | 0.002 | 0.5 | 2.0 |
9722-L53 | 9722-M53 | 50 | 0.004 | 0.8 | 4.0 |
9722-L56 | 9722-M56 | 75 | 0.008 | 1.6 | 8.0 |
9722-L59 | 9722-M59 | 100 | 0.015 | 3.0 | 15 |
9722-L62 | 9722-M62 | 150 | 0.035 | 7.0 | 35 |
9722-L65 | 9722-M65 | 200 | 0.1 | 20 | 100 |
9722-L68 | 9722-M68 | 300 | 0.25 | 50 | 250 |
9722-L71 | 9722-M71 | 350 | 0.5 | 100 | 500 |
9722-L74 | 9722-M74 | 400 | 1.2 | 240 | 1200 |
9722-L77 | 9722-M77 | 450 | 2.5 | 500 | 2500 |
9722-L80 | 9722-M80 | 500 | 8.0 | 1600 | 8000 |
9722-L83 | 9722-M83 | 600 | 20 | 4000 | 20000 |
† Viscosity ranges in mm²/s represent flow times of 250 to 1000 seconds for size 25
and 200 to 1000 seconds for all other sizes
and 200 to 1000 seconds for all other sizes
The Cannon-Ubbelohde Dilution Viscometer measures intrinsic viscosity of transparent Newtonian liquids according to ASTM D445 and ISO 3104. Minimum sample volume is 8 mL (can be diluted to 40 mL). Minimum bath depth is 305 mm (12 in). The CANNON CT-518 bath is recommended. (CT-500 and CT-600 baths are not deep enough.) Viscometer constant is the same at all temperatures. Both calibrated and uncalibrated viscometers are available. Calibrated viscometers include a certificate of calibration and instruction sheet. Holders are sold separately.